| A
"Check-List" for Measurement is as Follows:
1..
Always use "X 10" probes: they load the DUT (device-under-test) ~ 10 Meg
ohms @ ~ 10 pfd. A "X 1" probe offers 1 Meg ohm @ ~ 50 pfd. The designation
"X 10" refers to the attenuation of the signal by the probe (not gain).
In order to attain such light loading by the scope--while maintaining bandwidth--this
tradeoff is required.
2..
Make sure the probes are compensated (adjust trimmer at connector housing)
if attaching them to a different scope. This ensures maximum fidelity and
bandwidth of the signals being eyeballed.
3..
Use the shortest ground lead or clip-lead possible: the shorter the better!
Excessive ground lead length introduces unnecessary inductance and can
alter the displayed signal, as well as reducing the scope's effective bandwidth
(acts like a lowpass filter).
4..
When Measuring very high frequencies--especially in tight spaces--consider
using a RF probe (see figure). Also, there are--so-called--FET or active
probes, which are non-loading (almost) wideband probes with built-in amplifiers.
5..
When buying probes for your oscilloscope, make sure the probe is of sufficient
bandwidth for your particular scope: the probe is the first-order bandwidth
determinant of any scope.
Some
scopes have such a wide bandwidth, that no passive probe is able to do
it justice, and the only way to use the maximum bandwidth of this type
of scope is to drive the scope from a 50 ohm source through a 50 ohm coax,
terminated into 50 ohms at the scope's input. In fact, some high performance
scopes have a 1Meg ohm/50 ohm termination switch for just such occasions.
TOP
Input
Amplifier
TOP
Input
Multiplexer
Channel
A, Alternate, Chop, A-B, Channel B
TOP
Vertical
Amplifier, Phase Inverter
TOP
Vertical
Deflection Amplifier
TOP
Delay
Lines
A Delay
in Discussing a Loose End
Something
that cheap scopes are guilty of, is the fact that you never see the leading
edge of the event that triggered the sweep. That is to say, the sweep is
triggered by the leading edge of a pulse, and by the time the beam starts
across the screen it is now displaying the signal ~ 100 nsec after the
event. To solve this little "hand-is-quicker-than-the-eye" problem, the
Keebler
Elves added wideband analog delay lines between the vertical amplifier
and the CRT's deflection plates. So, by the time the beam has started its
short trip across Mr. CRT's face, the signal that triggered the sweep is
finally dribbling out of that long delay line.
TOP
Vertical
Deflection Plates
TOP
Triggered
Sweep / Timebase Generator
|
Trigger
Level & Trigger Polarity
|
 |
|
TOP
Sweep
Voltage: Horizontal Scan
On
the Level
Oscilloscopes
have several sync modes: internal sync; external sync; line sync and delayed
sync. The selection of any of these sync sources sends the signal to the
"sync detector." For example: if the internal sync is selected, it sends
the input signal, picked-off from the vertical input amplifier, to the
"sync detector" which is a fast analog comparator that gets its other comparison
input from a front panel mounted control, called the "LEVEL" control. This
allows the user to preset the exact voltage level and polarity, at which
the scope "triggers" the sweep generator to sweep the beam across the screen
once. Every time the signal goes above this preset voltage (except during
an active sweep time), a new sweep is started: this is called "Triggered
Sweep."
Oh,
By The Way...
There
is a sweep begun when ever the signal reaches the trigger level, regardless
of whether it is a rising or falling voltage. This means that--in the case
of a sine wave--the scope is just as apt to start the display at zero degree
phase, as 180 degree phase. To solve this little oversight, a switch is
used to tell the sweep generator which excursion--rising or falling--to
trigger on: this is called the "SLOPE." (+) SLOPE (-)
The
figures show several combinations of settings.
TOP
Horizontal
Deflection Amplifier
TOP
TOP
Horizontal
Deflection Plates
|